Cleaning and Maintenance
Cleaning
Console
Should the console need cleaning, wipe it down with a sterile cloth and mild
cleaning solution. If needed, wipe the console with a disinfectant.
Warning To avoid electric shock and potentially fatal injury,
unplug the Crossre2 console from the electrical outlet
before cleaning.
Do not sterilize the console or immerse it in any liquid.
Doing so will damage the unit.
Do not clean the console with alcohol, solvents, or
cleaning solutions that contain ammonia. Doing so will
damage the unit.
Footswitch
Consult the footswitch user guide for cleaning and reprocessing instructions.
RF Handpiece
RF handpieces are intended for single use only and should not be cleaned,
sterilized, or reused.
Shaver Handpiece
Consult the appropriate user guide for cleaning and reprocessing instructions.
Disposable attachments are intended for single use only and should not be
cleaned, sterilized, or reused.
Maintenance
The Crossre2 console requires no preventative or periodic maintenance.
However, Stryker recommends you reboot the system daily for best
performance.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals