Recommended Separation Distances Between Portable and Mobile RF Communications Equipment
and the Crossre2 System
The Crossre2 System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the Crossre2 System can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Crossre2 System as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power (W) of
transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 3.70
10 3.70 2.33 7.37
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection
from structures, objects, and people.
Symbol Glossary
This device and its labeling contain symbols that provide important
information for the safe and proper use of the device. These symbols are
dened below.
Warning Symbols
Warning/Caution:
See instructions for
use
Hazardous voltage present
Front Console Symbols
Power Select
Up Down