EasyManuals Logo

Stryker Crossfire 2 User Manual

Stryker Crossfire 2
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
EN-13
DRAFT
Setup and Device Connections
Stryker Endoscopy considers instructional training an integral part of the
Crossre2 system. Your Stryker Endoscopy sales representative will perform
at least one inservice at your convenience to help you set up your equipment
and instruct you and your sta on its operation and maintenance. Please
contact your local Stryker Endoscopy representative to schedule an in-service
after your equipment has arrived.
Warning Be sure that no liquid is present between connections
to the console and the handpiece. Connection of wet
accessories may lead to electric shock or electrical short.
To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
Use only hospital-grade power cables. Using other cables may result in
increased RF emissions or decreased immunity from such emissions.
Only the handpieces and disposable attachments are suitable for use in the
patient environment. The console and footswitch are not sterile devices and
should not enter the sterile eld.
The Crossre2 System is compatible only with the Stryker handpieces and
footswitches listed in this manual. Do not connect any equipment not
specied in this manual, as unexpected results or serious injury will occur.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals