Output Power (COAG) versus Load Resistance
Output Power (Coag) versus Load Resistance
0
10
20
30
40
50
60
70
80
90
100
0 100 200 300 400 500 600 700 800 900 1000
Load Resistance (Ohms)
Power (W)
Coag 1
Coag 2
Coag 3
Maximum Open Circuit Voltage versus Set Point
Maximum Output Voltage (RMS) versus Setting
0
50
100
150
200
250
300
350
400
Coag
1
Coag
2
Coag
3
1 2 3 4 5 6 7 8 9 10 11
Cut Level
Voltage (Vrms)