80 2131-209-001 REV A www.stryker.com
Return To Table of Contents
BED CLEANING
Hand wash all surfaces of the bed with warm water and mild detergent. Wipe the bed clean and DRY THOROUGHLY
to avoid build up of cleaning solution. Do not steam clean or hose off the bed. Do not immerse any part of the bed.
Some of the internal parts of the bed are electric and may be damaged by exposure to water.
Suggested cleaners for bed surfaces:
• Quaternary Cleaners (active ingredient - ammonium chloride).
• Phenolic Cleaners (active ingredient - o-phenylphenol).
• Chlorinated Bleach Solution (5.25% - less than 1 part bleach to 100 parts water).
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer’s guidelines
for proper disinfecting.
CAUTION
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF
USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures
must be taken to ensure the beds are wiped with a damp cloth soaked in clean water and thoroughly dried following
cleaning. Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly
causing premature corrosion of critical components. Failure to follow the above directions when using these types of
cleaners may void this product’s warranty.
MATTRESS CARE
Inspect the mattress after each use. Discontinue use if any cracks or rips are found in the mattress cover as these
openings may allow fluid to enter the mattress.
Cleaning