The IInnTToouucchh and IIssoolliibbrriiuumm system is not intended:
• for use with patients 35 inches or less in height
• for use with patients that weigh less than 50 pounds
• for use with a full oxygen tent
• to support more than one individual at a time
• for use with a patient less than two years old
• in a home healthcare environment
• as a sterile product
• for use in the presence of flammable anesthetic mixture with air rich in oxygen or nitrous oxide
NNoottee -- When IIssoolliibbrriiuumm is on the IInnTToouucchh bed, a half-oxygen tent within the sidreails is allowable. Do not enclose any
part of the support surface foot box with the half-oxygen tent.
IIssoolliibbrriiuumm is not intended for use on patients with the following:
• unstable fractures
• unstable spinal cord injuries
IIssoolliibbrriiuumm Lateral Rotation and Turn Assist features are for use only after you complete a clinical assessment. This is to
make sure that no adverse reaction would result in patients with the following:
• spinal cord injury
• skeletal traction
• significant hemoptysis
• bleeding disorders
• rib fractures
• for whom the techniques cause increased dyspnea or wheezing
• hemodynamic instable
• increased intracranial pressure
• skeletal traction
SSppeecciiffiiccaattiioonnss
Model 2971/2972 IIssoolliibbrriiuumm
Length 84 in. 213,4 cm
Width 35 in. 88,9 cm
Thickness 8 - 10.5 in. 20.3 - 26.7 cm
Weight 90 lb 40.8 kg
Therapeutic weight 50 lb to 350 lb 22.7 kg to 158.7 kg
Safe working load IIssoolliibbrriiuumm
(Sum of the patient and accessory
weight)
460 lb 208.6 kg
EN 4 2971-009-002 Rev D.0