38
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
L10 AIM Light Source is intended for use in the electromagnetic environment specied below.
The customer or the user of L10 AIM Light Source should ensure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment:
Guidance
Electrostatic Discharge
(ESD)
IEC61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floors should be
wood, concrete, or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast transient/
burst
IEC61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
(if applicable)
±2kV line to ground
±1kV for input/output
lines
(if applicable)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential mode
±2kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC61000-4-11
0% U
T
0.5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 sec.
0% U
T
0.5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 sec.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of L10 AIM
Light Source requires
continued operation
during power mains
interruptions, it is
recommended that
L10 AIM Light Source
be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60Hz) magnetic eld
IEC 61000-4-8
30 A/m 30 A/m Power-frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: U