All fasteners secure.
Siderails move and latch properly.
All casters lock with brake pedal engaged.
All casters secure and swiveling properly.
Fowler operates and latches properly.
Trendelenburg/Reverse Trendelenburg operating properly.
Ground chain intact.
No leaks at hydraulic connections.
Hydraulic jacks holding properly.
Hydraulic oil level sufficient.
Body restraints working properly.
I.V. pole intact and operating properly.
Oxygen bottle holder intact and operating properly.
No rips or cracks in mattress cover.
Engage the drive wheel and ensure it is operating properly.
No excessive play in the drive handles.
Press the handle switches. Unit should not move unless the handles are pushed forward or pulled back.
Press the handle switches. Move the handles forward and back and verify the unit responds properly.
Confirm battery powered functionality.
No cables worn, pinched or frayed.
All electrical connections tight.
All grounds secure to the frame.
Ground impedance not more than 100 milliohms max: Test point(s) include electronics enclosure and motor
chassis mounted to base of unit.
Current leakage not more than 300 microamperes (per UL 60601-1).
Batteries sufficiently charged (optional scale system).
Display housing intact and not damaged (optional scale system).
Display label intact and not damaged (optional scale system).
Load cells intact and not damaged (optional scale system).
Scale calibrated properly. Recalibrate, if necessary (optional scale system).
CHECKLIST
Stretcher Serial Number:
Completed by: _______________________________________ Date: _________________
Note
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
Preventative Maintenance