4. Inspect the rover. See
To Inspect the Equipment
(page 51).
5. Refill the manifold holder with new, unused manifolds, as required. See
For Use With
(page 12) for ordering information.
6. If the rover will be used again, use the rover handle to push and relocate the rover. See
To Setup the Rover
(page 36). If
the rover will no longer be used, use the rover handle to push and relocate the rover to a storage area. See
Storage and
Handling
(page 58).
55..22.. TToo IInnssppeecctt tthhee EEqquuiippmmeenntt
NNoottee -- Only biomedical equipment technicians trained and experienced in the maintenance of this reusable medical
device should inspect and maintain this equipment.
WWAARRNNIINNGG -- Upon initial receipt and before each use, inspect each component for damage. Do not use the product if
damage is apparent.
WWAARRNNIINNGG -- Always comply with the inspection interval to ensure the safe and effective use of the equipment.
WWAARRNNIINNGG -- Do not disassemble, modify, or repair this product without the authorization of the manufacturer. Contact
Stryker for service.
NNoottee -- For service, contact your Stryker sales representative or call service. See
Contact Information
(page 4). Outside
the US, contact your nearest Stryker subsidiary.
NNoottee -- Maintenance documentation for this equipment is available upon request to Stryker-authorized service
personnel only.
NNoottee -- Routine and careful inspection of the equipment is the best method for determining the service life of the
equipment. See Inspection Criteria and Actions.
TTaabbllee 2299 –– IInnssppeeccttiioonn CCrriitteerriiaa aanndd AAccttiioonnss
IInntteerrvvaall CCrriitteerriiaa AAccttiioonn
Before initial use Inspect the equipment for damage or
missing components.
If damage is apparent, replace the
equipment. See
For Use With
(page 12)
section.
Make sure the rover and docker operate
properly.
See
To Dock the Rover
(page 54) and
To
Test the Rover
(page 32).
Before each use and after each
cleaning and disinfection.
Inspect the equipment for damage or
missing components.
If damage is apparent, replace the
equipment. See
For Use With
(page 12).
Inspect the equipment for corrosion,
discoloration, pitting, cracked materials,
or unacceptable deterioration on any
external surfaces, including labels and
product markings.
Inspect the canister exterior, smoke
evacuator filter cover, and infrared
communication window for cracks or
damage.
Inspect the caster and make sure the
locks function properly.
Inspect the power cord for cuts and the
power cord plug for bent pins.
700000969828 Rev-AA 51 EN