66.. RReeffeerreenncceess
66..11.. SSttoorraaggee aanndd HHaannddlliinngg
CCAAUUTTIIOONN -- Always store and transport the equipment within the specified environmental condition values throughout its
useful life. See
Specifications
(page 66).
CCAAUUTTIIOONN -- Always call service before transporting or storing this equipment in freezing conditions. See
Contact
Information
(page 4). Failure to comply will cause the expansion of frozen internal fluid to damage the equipment.
NNoottee -- The rover does not have to be connected to facility power when not in use.
66..22.. TTrroouubblleesshhoooottiinngg
WWAARRNNIINNGG -- Do not disassemble, modify, or repair this product without the authorization of the manufacturer. Contact
Stryker for service.
NNoottee -- For service, contact your Stryker sales representative or call service. See
Contact Information
(page 4). Outside
the US, contact your nearest Stryker subsidiary.
TTaabbllee 3311 –– RRoovveerr OOppeerraattiioonn
OOBBSSEERRVVAATTIIOONN CCOORRRREECCTTIIVVEE AACCTTIIOONN
The rover does not power up and the power switch is in the
ON position.
Power cord is not connected or is not connected securely.
Connect the supplied power cord or make sure the power
cord is connected securely. See .
Facility power is not available or off. Make sure facility
power is provided to the hospital-grade power receptacle.
The rover does not recognize a manifold. Manifold is altered or damaged. Replace the manifold. See
For Use With
(page 12). See
To Install Disposable
Components
(page 37).
No vacuum pump action after the START SUCTION button
is touched.
Canister is full and an error occurs. Dock the rover. See
To
Dock the Rover
(page 54).
The rover is damaged. Remove the rover from use.
Contact service. See
Contact Information
(page 4).
The rover’s fluid suction is weak or insufficient. Manifold is not installed correctly. Make sure the manifold
is seated and locked in place. See
To Install Disposable
Components
(page 37).
Unused manifold ports are open. Close all unused inlet
ports of the manifold installed in the canister.
Suction tubing connection is not secure. Make sure all
suction tubing connections are secure.
Unused suction tubing is not clamped. Clamp any suction
tubing not in use.
Suction tubing is blocked or damaged. Clear or replace any
suction tubing in use.
Suction accessory is blocked or damaged. Clear or replace
the suction accessory.