44..33.. TToo OOppeerraattee tthhee RRoovveerr
WWAARRNNIINNGG
HHIIGGHH SSUUCCTTIIOONN HHAAZZAARRDD [[MMAAXX == 552200 mmmm--HHgg]]
The effectiveness of aspiration is dependent upon the intensity of the suction applied.
Always use the minimum suction limit range required to achieve the desired clinical outcome.
Always follow your local healthcare guidelines for suction limits.
The suction limit range of the collection canister may be adjusted by using the manual suction control dial on the main
control panel. Interruption and restoration of rover power while suction is ON, whether accidental or intentional, does not
reset the suction limit range to zero. See BS EN ISO 10079-1: 2015 + A1: 2019, clause 9.11. Use caution when
activating suction with a high suction limit range.
Death or serious injury can result from the improper application of suction levels.
WWAARRNNIINNGG
IINNAAPPPPRROOPPRRIIAATTEE CCOONNNNEECCTTIIOONN HHAAZZAARRDD
Do not connect directly to chest tubes.
Do not connect to closed wound drains.
Do not connect directly to tracheal tubes.
Not for use as a suction source for: Intermittent suction applications, Patient positioner devices, and Organ stabilizer/
positioner devices.
Death or serious injury can result from inappropriate connections.
WWAARRNNIINNGG -- Always operate the product within the specified environmental condition values.
44..33..11.. TToo MMaannaaggee FFlluuiidd SSuuccttiioonn
WWAARRNNIINNGG
FFLLUUIIDD WWAASSTTEE LLEEAAKKAAGGEE HHAAZZAARRDD
Always make sure rover power is ON when collecting fluid waste. The rover can only detect a full canister if the rover
power is ON. If rover power is OFF, fluid waste leakage can occur.
WWAARRNNIINNGG -- The collection canister scale and main control panel display are not diagnostic tools. Do not use the scale
or the display to determine the amount of fluid lost from or retained by the patient.
NNoottee -- Make sure the rover has been properly prepared for operation. See
To Setup the Rover
(page 36).
NNoottee -- The rover provides an adjustable suction limit of 50 to 520 mm-Hg measured with all manifold inlet ports closed.
NNoottee -- To reset the fluid level value to zero, touch the RESET VOLUME button on the main control panel display. The
canister fluid level value will reset to a zero value.
EN 40 700000969828 Rev-AA
DRAFT NOT FOR SALE