66..88.. EElleeccttrroommaaggnneettiicc CCoommppaattiibbiilliittyy
product safety to verify accuracy of this information
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc eemmiissssiioonnss
The Neptune S Rover is intended for use in the electromagnetic environment specified below. The customer or the user
of the Neptune S Rover should assure that it is used in such an environment.
EEmmiissssiioonnss tteesstt CCoommpplliiaannccee EElleeccttrroommaaggnneettiicc eennvviirroonnmmeenntt -- gguuiiddaannccee
RF emissions
CISPR 11
Group 1 Neptune S Rover uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A Neptune S Rover is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
WWAARRNNIINNGG -- This equipment/system is intended for use by
healthcare professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment.
Mitigation measures may be necessary, such as reorienting or
relocating the Neptune S Rover or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
This Neptune S Rover is intended for use in the electromagnetic environment specified below. The customer or the user
of the Neptune S Rover should assure that it is used in such an environment.
IImmmmuunniittyy tteesstt IIEECC 6600660011 tteesstt lleevveell CCoommpplliiaannccee lleevveell EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt -- gguuiiddaannccee
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 20% ??.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV at 100 kHz repetition
frequency for power supply
lines
±1 kV at 100 kHz repetition
frequency for input/output
lines
± 2 kV at 100 kHz repetition
frequency for power supply
lines
±1 kV at 100 kHz repetition
frequency for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV line(s) to
line(s)
± 0.5 kV, ± 1 kV, ± 2 kV line
(s) to earth
± 0.5 kV, ± 1 kV line(s) to
line(s)
± 0.5 kV, ± 1 kV, ± 2 kV line
(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.