EasyManuals Logo

Stryker Neptune S User Manual

Stryker Neptune S
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
IInntteerrvvaall CCrriitteerriiaa AAccttiioonn
Inspect the power cord receptacle for
bent pins or bent contacts.
Inspect the infrared communication
window for any dirt or debris.
Thoroughly remove any dirt or debris on
the infrared communication window. See
To Clean and Disinfect the Rover
(page
50).
Six months Inspect the replacement date on the label
of the fluid suction HEPA filter. The fluid
suction filter life is 500 hours.
Replace the fluid suction HEPA filter
every six months or as indicated on the
main control panel display. See
For Use
With
(page 12) and
To Replace the HEPA
Filter
(page 52).
NNoottee -- If any component must be discarded, see
Disposal/Recycle
(page 65).
55..33.. TToo RReeppllaaccee tthhee HHEEPPAA FFiilltteerr
Failure to comply may cause infection and result in healthcare staff injury.
WWAARRNNIINNGG -- Do not touch the filter media. The media filters the air evacuated from fluid collection canister of the rover
before the air is vented. Failure to comply may cause media damage and result in patient and/or healthcare staff injury.
FFiigguurree 4400 PPrreevveenntt DDaammaaggee ttoo FFiilltteerr MMeeddiiaa
EN 52 700000969828 Rev-AA
DRAFT NOT FOR SALE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Neptune S and is the answer not in the manual?

Stryker Neptune S Specifications

General IconGeneral
BrandStryker
ModelNeptune S
CategoryMedical Equipment
LanguageEnglish

Related product manuals