IItteemm IInnssppeecctt
Fowler bumper and screws
Base All welds are intact, not cracked, or broken
Wheels Check and adjust wheel locks
Retractable head section oxygen bottle holder (optional) Straps and clips for wear
Defibrillator platform (optional) Straps not frayed or torn
Latch hooks are intact and secure
FFoooott eenndd ffaasstteenneerr ppaarrtt rreeppllaacceemmeenntt sscchheedduullee
For PPeerrffoorrmmaannccee--LLOOAADD compatible cots, you must replace the foot end fastener parts every 18,078 calls. This is to make
sure that the PPeerrffoorrmmaannccee--LLOOAADD remains functional. Follow this call volume time table to remain compliant with this
requirement.
CCaallllss ppeerr ddaayy MMoonntthhss
7 Not applicable
8 77
9 67
10 59