NNoottee -- Make sure that the X-frame guard screw hole on the new inner tube is facing the top of the cot when
reassembling.
13.Verify proper operation before you return the product to service.
SSiiddeerraaiill aasssseemmbbllyy rreeppllaacceemmeenntt ((ssttaannddaarrdd))
TToooollss rreeqquuiirreedd::
T25 Torx driver
PPrroocceedduurree::
1. Raise the product to the highest height position.
2. Raise the siderail to the up and locked position.
3. Using a T25 Torx driver, remove the three spindle screws that secure the siderail assembly.
4. Remove the siderail.
5. Reverse steps to install the new siderail assembly.
6. Verify proper operation before you return the product to service.
SSiiddeerraaiill aasssseemmbbllyy rreeppllaacceemmeenntt ((XXPPSS ooppttiioonn))
TToooollss rreeqquuiirreedd::
T25 Torx driver
1/4'' hex wrench
3/16'' hex wrench
Slotted screwdriver
Torque wrench (ft-lb)
Rubber mallet
PPrroocceedduurree::
1. Raise the product to the highest height position.
2. Remove the mattress.
3. Using a T25 Torx driver, remove the button head cap screw (A) and black bumper (B) on the side where you are
replacing the siderail (Figure 15). Save the screw and bumper.
6086-009-002 Rev AB.2 21 EN

Table of Contents

Other manuals for Stryker Performance-PRO XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Performance-PRO XT and is the answer not in the manual?

Stryker Performance-PRO XT Specifications

General IconGeneral
BrandStryker
ModelPerformance-PRO XT
CategoryMedical Equipment
LanguageEnglish

Related product manuals