12
8.1 Changing the settings
The Settings menu is populated with various settings, some of which have default options which can
be changed to suit your individual preferences and geographical location. To change a setting, touch
to select a setting. A pop-up screen will appear. Select the required option from the pop-up screen
then touch the main screen behind or back key to exit.
Device settings
Language: Select either ‘English’ or ‘French’.
Device storage: Displays the total amount of internal memory used and the amount of
memory remaining.
Camera gain: Select either ‘Low’ (default) or ‘High’.
The ‘High’ setting will increase the brightness of the image compared to the
‘Low’ setting. The ‘Low’ setting should be suitable for most melting point
determinations.
Error reporting: Displays details of any errors in QR code format which can be sent to Bibby
Scientific Ltd. for diagnostic purposes.
Notifications: Can be set such that the unit sends a notification for the following:
❖ Measurement completed.
❖ Calibration reminder can be set to Never (default), Weekly,
Monthly, Bi-annually or Annually.
Date & time: Can be set for regional time zones. Set the date and local time before first
use.
Check USB for Allows you to perform software updates from a USB flash drive. Updates
an update: can be downloaded from www.stuart-equipment.com or contact
techsupport@bibby-scientific.com. See section 18 for how to perform a
software update.
About: Displays information about the currently installed version of software.
Display formats
Temperature units: Select either ‘°Celsius’ or ‘°Fahrenheit’.
Time format: Select either ‘12 Hour’ or ‘24 Hour’.
Results
Save the melt video Select ‘ON’ to save the melting video for future reference. Use the slider
with the report: button to turn on or off.
Encrypt all results Select ‘ON’ to encrypt the results files to prevent post-run editing. Use the
files: slider button to turn on or off.
Note: The SMP50 can serve as part of a US FDA 21 CFR part 11 compliant system relating to its use
to generated results in an electronic format. Results generated on the instrument can be copied to a
USB flash drive in an encrypted format which is protected against tamper or change. The encrypted