CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018 66 of 112 DLM 97200-03 EN Rev. A
Modbus Register Table
The reference addresses in the table below are the values sent over the Ethernet connection.
Function Code
0x02
Discrete Inputs (read only)
Status 1000 5 bits
New Data (0) // bit 0
Error (1) // bit 1
Warning (2) // bit 2
Standby (3) // bit 3
Alarm1 (4) // bit 4
Alarm2 (5) // bit 5
Coils (read) 0x01
State 2000 2 bits
Run (0) // bit 0
Stop (1) // bit 1
Coils (write) 0x05
Run 2000
Stop 2001
Input Registers (read only) 0x04
Instrument Family 3010 // 16-bit unsigned
Instrument Model 3011 // 16-bit unsigned
Serial Number 3012 // 16-bit unsigned
Reading Time – Year 3100 // 16-bit unsigned
Reading Time – Month 3101 // 16-bit unsigned
Reading Time – Day 3102 // 16-bit unsigned
Reading Time – Hour 3103 // 16-bit unsigned
Reading Time – Minute 3104 // 16-bit unsigned
Reading Time – Second 3105 // 16-bit unsigned
Error Type 3200 // 16-bit unsigned
Error Time – Year 3201 // 16-bit unsigned
Error Time – Month 3202 // 16-bit unsigned
Error Time – Day 3203 // 16-bit unsigned
Error Time – Hour 3204 // 16-bit unsigned
Error Time – Minute 3205 // 16-bit unsigned
Error Time – Second 3206 // 16-bit unsigned
TOC Value IEEE 32-bit float
TOC Value Low Word 3300 // 16-bit unsigned
TOC Value High Word 3301 // 16-bit unsigned
TOC Units 3302 // 16-bit unsigned
1 = ppt
2 = ppb
3 = ppm
TOC Trend Value IEEE 32-bit float
TOC Trend Value Low Word 3400 // 16-bit unsigned
TOC Trend Value High Word 3401 // 16-bit unsigned
TOC Trend Units 3402 // 16-bit unsigned
1 = ppt/hr
2 = ppb/hr
3 = ppm/hr