CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018 93 of 112 DLM 97200-03 EN Rev. A
Cleaning the Sensor
You can clean the external housing of the Sensor with a damp cloth using water or non-abrasive
cleaners. Turn off power to the Sensor and disconnect it from main power or remove the battery prior
to cleaning. Do not spray liquids directly on the Sensor. Wipe dry with a clean, soft cloth.
The screen may be cleaned with a soft cloth or cotton pad. Isopropyl Alcohol may be used, but insure
that all solvent residue is removed. Clean any condensation or moisture from any source
immediately. If the touch screen becomes fractured or cracked, do not touch it and call technical
support.
Cleaning the Vial Sampling Kit
The vial sampling kit should be rinsed before and after use to remove residue from samples or
standards that may remain on the wetted parts. This can easily be performed by running grab
samples through the sampler from a vial of DI water for 15 minutes or until sample readings stabilize.
Table 5: CheckPoint 2-Year Maintenance Worksheet