76 A-96.250.571 / 050517
AMI Codes-II
Program List and Explanations
2.2.1.5 Ver. History: Shows the verification values of the last verifications:
Number: Verification counter.
Date, Time: Date and time of the verification.
Absorbance: Measured absorbance of the reference kit.
Reference value: True value of the reference kit according to label.
2.2.2 pH Electrode: Only available if option pH is installed.
Current Value: Shows the actually measured pH value.
Raw value: Shows the actual electrode voltage in mV.
2.2.2.5 Cal. History: Shows the calibration values of the last pH sensor
calibrations. Offset in mV and slope in mV/pH.
Number: Calibration counter.
Date, Time: Date and time of the calibration.
Offset: Vertical shift above or below the zero point.
Slope: Characteristic of the pH sensor expressed in mV/pH.The
calculated slope is used for the entire measuring range of the
electrode.
2.2.3 Miscellaneous:
2.2.3.1 Case Temp: Shows the current temperature in [°C] inside the
transmitter.
Typical offset of pH electrode: < +/- 30 mV.
Max. tolerated offset: < +/- 60 mV
Typical slope of pH electrode: 5565 mV/pH unit.
Max. limits: 4065 mV/pH