TGA 2950 4–19
Status Codes
Table 4.1
(continued)
Code Meaning
Hot The temperature is
beyond the set point, and
the instrument cannot
remove heat fast enough
to follow the thermal
program. This is usually
caused by a large ballistic
jump to a lower tempera-
ture.
Initial The temperature is being
equilibrated to the desired
set point. When the
temperature has reached
equilibrium, the status
will change to Ready.
Iso The thermal program is
holding the current tem-
perature isothermally.
Jumping The heater is jumping
ballistically to the set
point temperature.
Load The TGA is loading a
sample pan onto the
balance.
No Power No power is being applied
to the heater. Check the
heater switch and fuse.
(table continued)