Appendix D
TA I
NSTRUMENTS
TGA 2950D–8
Figure D.2
Removing the Furnace
Carriage Mounting Screws
7. Use the 3/32-inch ball driver supplied with
the EGA installation kit to replace the
screws with the new screws, spacers and
cleats provided with the kit. Refer to Figure
D.3. Replace the upper screws using the
long screws with spacers and replace the
lower screws using the short screws without
spacers.
Water
Line
(CCC)
Water Line (AAA)
Water Line
(CCC)
Mounting
Screws