oCare
TM
Pro 100
48
CAUTION:
1 In order to get fast and accurate results,
Make sure the device sit on your wrist skin, slightly
above the wrist bone (ulnar styloid processus), as
illustrated in Figure 8.
Keep the device-worn lower arm horizontal and still
while the device is taking the measurement.
Do not perform the measurement while the blood
flow is restricted (ex. taking blood pressure).
2 When PR/SpO
2
high (or low) limit is exceeded,
Medium priority alarm is activated (refer to Table 4
for details). Address the patient condition
immediately.
The PR/SpO
2
number on the display turns YELLOW.
The device takes the measurement continually until
the medium alarm condition is resolved.
3 When PR/SpO
2
high (or low) limit is changed from
default, a decimal point (.) is displayed after the reading as
a reminder.
4 If PR/SpO
2
results do not show in designated time, the low
priority alarm is triggered and the symbol is displayed
(in the lower left corner of the display) to indicate the
situation. Please adjust the watch position (slightly above
the wrist bone or ulnar styloid processus) for better signal