reading. If the situation persists, please refer to
“Troubleshooting” section for instruction.
5 When technical alarm is triggered, low priority alarm is
activated (refer to Table 4 for details). Address the device
condition immediately.
6 Tattoo or hairy hairs in the sensor site would interfere with
the measurement. Please switch hand or shave off the hairs
on the sensor site to ensure the best measurement
accuracy.
NOTE:
1 The device measures, collects, and stores up to 72 hours of
SpO
2
and pulse rate data with a 1-second data collection
rate.
2 When the memory is full, the device overwrites the oldest
existing data with the new data.
3 Syncing data in memory does not clear memory.
4 During normal measurement conditions the averaging time
is 8 seconds. During conditions such as those caused by
low perfusion, interference, the algorithm automatically
extends the amount of data required beyond 8 seconds.
5 If the signal from the sensor is inadequate, the last
measured SpO
2
and pulse rate values freeze for 25
seconds, and are then replaced with dashes.