EasyManua.ls Logo

TE RBK-X1 - Manual Calibration

TE RBK-X1
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RBK-X1 / RBK-X1C
Rev. B Jun. 2021
27
3.7 Manual Calibration
1.
RBK-X1 Set
RBK processor calibration temperature and cycle time. Factory setting: 500°C @ 15S.
2.
RBK-X1 Current
RBK processor real temperature and cycle time. (When calibration cycle start, cycle time will count down from 15S to 0S.)
Manual Calibration Procedure
1. Inactivate the Centering in HMI Maintenance page.
2. Enter Manual Calibration page, allow processor to stabilize at least 25 minutes.
3. Connect UHI probe to a calibrated meter.
4. Carry out 3 calibration readings (cool UHI to 23°C± C between each cycle).
Calculate AVERAGE VALUE.
5. Compare measured value (average) with desired value of 13C ± C.
New OFFSET= old OFFSET” + (“AVERAGE VALUE - 130°C) x 2
6. Wait 25 minutes for temperature to stabilize.
7. Repeat 3 (and 4 if necessary).
8. After OFFSET has been adjusted correctly it is necessary to wait 15 minutes for temperature to stabilize
prior to normal use.
It is recommended to carry out an auto calibration after manual calibration in order to remove the
Cal message in the display panel.
Note: If this does not disappear it is necessary to replace the heating element and do auto-calibration.
1
2

Table of Contents