Now the device is ready for measurements on
solid color patches. In this case a yellow patch
was measured.
Density trend: It is recommended to increase
the density by 0.11.
Color deviation: The color distance to the ref-
erence is 4.8 ∆E*. After following the density
trend it can be as low as approx. 1.4 ∆E*.
A closer color match will be limited because
the paper and ink characteristics do not allow
to get closer to the reference standard.
Please note that the density values are ob-
tained without the polarizing lter activated
and therefore are lower as values obtained
with polarizing lter. Please ensure that the
right measurement conditions are set.