Performance Characteristics
Table 6-2 (cont)
Option
01
(DMM) Electrical Characteristics
Characteristics
Performance Requirements
CONTINUITY
Response
Time
Approximately
0.1
second.
Threshold Resistance
1011
±1
11.
TEMPERATURE
Accuracy
+18°C
to
+28°C
Ambient
Temperature
±(2%
of
reading
+ 1.5°C).
-15°C
to
+18°C
and
+28°C
±(2%
of
reading
+ 2.0°C).
to
+55°C
Ambient Temperature
Probe
Tip Measurement
Range
-62°C
to
+230°C
in
one
range.
Resolution
0.1
°C or
0.1
of.
ADDITIONAL CHARACTERISTICS
Warmup time to Meet Electrical
Specification
Maximum
Voltage between Inputs
from either Input
to
Ground
DC
to
20
kHz
Above
20
kHz
45
minutes.
500 V
rms;
700 V peak.
10
7
V·Hz.
NOTE
For
AMPS
modes, maximum voltage between inputs is limited
by
maximum
input current.
2465A/2455A/2445A Operators