EasyManuals Logo
Home>Tektronix>Test Equipment>MSO58

Tektronix MSO58 Printable Help

Tektronix MSO58
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #460 background imageLoading...
Page #460 background image
Flattop2 window
This window has the lowest scallop loss of any of the windows. It also has a wider resolution bandwidth but lower side lobe
attenuation. Also, it is unique because the time domain shape has negative values.
NOTE. The Flattop2 window is useful for high accuracy magnitude measurements for signals that do not require very narrow
bandwidth.
Gaussian window
This is the default window function (see the next figure). It is unique in that the time-domain shape of an exponential Gaussian
function transforms into a Gaussian exponential shape in the frequency domain.
NOTE. This window provides optimal localization in both the time and the frequency domain.
Measurement concepts
440 MSO54, MSO56, MSO58, MSO58LP, MSO64 Help

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals