Self Calib
ration. The self calibration routine optimizes the oscilloscope accuracy
for the ambient temperature. For maximum accuracy, perform a self cal if the
ambient temperature changes by 5 °C (9 °F ) or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up.
Follow the directions on the screen.
Factory calibration uses externally-generated voltages, and requires specialized
equipment. The recommended interval is one year. See Contacting Tektronix
on the copyright page for information on having Tektronix perform a Factory
Calibr
ation of your oscilloscope.
File Utilities f or the USB
Flash Drive
Onefolderisalwaysdesignatedasthecurrent folder. The current folder is the
default location to save and recall files.
You c an use the File Utilities menu to do the following tasks:
List
the contents of the current folder
Select a file or folder
Navigate to other folders
Create, rename, and delete files and folders
Format a USB flash drive
O
ptions
C
omments
N
avigates to the selected USB flash drive folder. Use the
multipurpose knob to select a file or folder, then select the
Change Folder menu option.
C
hange Folder
To return to the previous folder, select the ↑Up folder item and
then select the Change Folder menu option.
New Folder
Creates a new folder in the current folder location, named
NEW_FOL, and displays the Rename m enu for changing the
default folder name.
Rename (filename or
folder)
Displays the Rename screen to rename a folder or file, described
next.
Delete (filename or
folder)
Deletes the selected file name or folder; a folder must be empty
before you can delete it.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 103