EasyManua.ls Logo

Tektronix TDS2000C series - Page 133

Tektronix TDS2000C series
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix A: Specications
Table 3: Vertical Specications (cont.)
Characteris
tic
Description
V/div values are accurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
TDS2001C TDS2002C, 2004C TDS1012C-EDU,
TDS2012C, 2014C
TDS2022C, 2024C
DC to >50 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limi t at full.
<5 mV/div settings are limited
to 20 MHz BW
DC to >70 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limit at full.
<5 mV/div settings are
limitedto20MHzBW
DC to >100 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limit at full.
<5 mV/div settings are limited
to 20 MHz BW
DC to >200 MHz for 5 mV/div
through 5 V/div settings with
bandwidth
limit at full with
temperature between 0 and
35 °C.
DC to >160
MHz from
5 mV/div through 5 V/div
settings with bandwidth
limit at
full for temperatures
between 0 and 50 °C. <
5mV/div settings are limited
to 20 MH
zBW
TDS1001C-EDU TDS1002C-EDU
Analog Bandwidth, DC
Coupled, Sa
mple or
Average
DC to >40 MHz for 5 mV/div
through 5 V/div settings
with b
andwidth limit at full.
<5 mV/div settings are limited
to 20 M Hz BW.
DC to >60 MHz for 5 mV/div
through 5 V/div settings
with b
andwidth limit at full.
<5 mV/div settings are
limitedto20MHzBW
The Analog Bandwidth when the instrument is DC coupled. V/div values are accurate for probe attenuation settings of 1X. No
probe should be installed for these measurements.
TDS1001C-EDU, TDS2001C TDS1002C-EDU,
TDS2002C, 2004C
TDS1012C-EDU, TDS2012C, 2014C, 2022C, 2024C
Analog Bandwidth, DC
Coupled, Peak Detect
DC to >30 MHz for 5 mV/div
through 5 V/div settings
wit
h bandwidth limit at full.
Setting less than 5 mV/div
are limited to 20 MHz BW
DC to >50 MHz for 5 mV/div
through 5 V/div settings
wit
h bandwidth limit at full.
Setting less than 5 mV/div
are limited to 20 MHz BW
DC to >75 MHz for 5 mV/div through 5 V/div s ettings with
bandwidth limit at full. Setting less than 5 m V/div are limited
to 2
0MHzBW
An
alog Bandwidth
Selectio ns
20 MHz BW Limit ON/OFF
Lo
wer Frequency Limit, AC
Coupled
10 Hz
1 Hz when 10X passive probes are used.
Rise time is generally calculated from the following formula: Rise time in ns = 350 / Bandwidth in MHz
TDS2001C TDS2002C, 2004C TDS1012C-EDU,
TDS2012C, 2014C
TDS2022C, 2024C
= 7.0 ns = 5.0 ns = 3.5 ns = 2.1 ns
TDS1001C-EDU TDS1002C-EDU
Rise Time, typical
=8.8ns =5.8ns
The capability of the instrument to capture single event pulses using Peak Detect Acquisition Mode.
The minimum single pulse widths for guaranteed 50% or greater amplitude capture are as follows:
Sec/Div Setting
Minimum Pulse Width
50 s/div to 5 us/div
TDS1002C-EDU, TDS1012C-EDU, TDS2002C, TDS2004C,
TDS2012C, TDS2014C, TDS2022C, TDS2024C, TDS2001C
12 ns
Peak Detect Mode Pulse
Response
TDS1001C-EDU, TDS2001C
13 ns
DC Gain Accuracy, Sample
or Average Acquisition
±3%, 5 V/div through 10 mV/div.
±4%, 5 mV/div and 2 mV/div.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 109