Specifications
1-8
TPS2000 Series Digital Storage Oscilloscopes S ervice Manual
Table 1- 5: Oscilloscope environm ental characteristics
Temperature Operating
2
Nonoperating
2
p
0 °Cto50°C(32°F to 122 °F) -- 4 0 °Cto+71°C(--40°F to 159.8 °F)
Humidity Operating
3
Nonoperating
3
y
60% RH: 30 °Cto50°C(86°F to 122 °F) 60% RH: 55 °Cto71°C (131 °F to 160 °F)
90% RH
4
:0°Cto30°C(32°Fto86°F)
Cooling Method Forced air, temperature controlled.
Altitude, Operating and
Nonoperating
3,000 m (10,000 ft)
Random Vibrati on Operating Nonoperating
0.31 g
RMS
from 5 Hz to 500 Hz, 10 minutes on
each axis.
2.46 g
RMS
from 5 Hz to 500 Hz, 10 minutes on
each axis.
Mechanical Shock,
operating
1
50 g, 11 ms, half sine
1
With one battery pack installed.
2
Charge the battery packs within an ambient temperature range of 0 °Cto45°C(32°Fto113°F). Charging battery packs
outside of this range can damage cells or cause them to leak.
3
The battery packs are rated to operate between - 10 °C and +50 °C(-14°F and +122 °F) with less than 80% relative
humidity. Operating outside of this range can cause damage. Battery discharge capacity drops significantly at tempera-
tures below 0 °C(32°F) and above 45 °C(113°F).
4
Wet bulb.