TIP: Figure 26 on page 30 contains a flow chart for the measurement process.
The chart shows all of the functions that are associated with the
measurement routine and how to access them. Figure 26 references
Figure 27 for the measurement sequence as shown on page 31.
To start the measurement process, follow the steps below:
1. Select Measurement Mode & Core Diameter Size
a. From the main menu (Figure 4), enter into the Setup menu.
b. Select Mode and select the preferred mode of measurement.
c. Select Core diameter then choose the core diameter size. Select “None” if you
wish to measure rock samples.
d. Select “Back to menu” to go back to main menu
2. Select Measure
Ensure the meter is on the start measure screen that is displayed in Figure 21. Before proceeding
with the measurements, ensure that the meter is position in a free air space void of all metallic
objects.
Step 1: With the KT-10 v2 in free air, use SBP▲ to start the measurement process.
After about 1 second you will hear a short sound indicating the free air
measurements are complete.
Step 2: Immediately place the KT-10 v2 on the sample’s surface then use SBP▲.
When the reading on the sample is complete you will hear a sound; this sound is
different then the one heard during the free air measurements.
Step 3: Then immediately position the KT-10 v2 in free air once again for the final free
air measurements. Wait for the final sound, which will be the same as the first
tone heard in Step 1. This sound will indicate the final free air measurements are
complete and the reading(s) will be displayed on screen.
Note: A button icon will appear in the notification area confirming a button press.
Tip: To repeat the measurement process without saving the results, position the
unit in free air and go directly to Step 1.