3 Technical data
14
3 Technical data
3.1 WiFi data loggers
Measurement-specific data
The humidity sensor attains the highest degree of accuracy in
temperatures between + 5 °C and + 60 °C and 20% to 80% RH. If the
instrument is exposed to higher humidity for a long period of time, this
can falsify the readings by up to 3% RH. After 48 hours at 50% RH ±
10 % and +20 °C ± 5 °C, the sensor regenerates by itself.
CAUTION
Damage to the humidity probe
- The probe must never be exposed to a humidity level of 100 % RH for
longer than 3 days.
WiFi data loggers testo 160 TH testo 160 THE testo 160 E
Order number 0572 2021 0572 2023 0572 2022
Temperature measurement
Measuring range
-10 °C to 50 °C
see ext. probes
Accuracy
± 0.5 °C
Resolution
0.1 °C
Humidity measurement
Measuring range 0 to 100% RH (non-condensing) see ext. probes
Accuracy ± 2% RH @ 25 °C & 20 to 80% RH
± 3% RH @ 25 °C & <20% RH &
>80% RH
± 1% RH hysteresis
± 1% RH/year drift
Resolution 0.1% RH
Lux measurement
Measuring range see ext. probes see ext. probes
Accuracy
Resolution
UV measurement
Measuring range
see ext. probes
see ext. probes
Accuracy
Resolution