4 Operation
21
4 Operation
4.1 Commissioning
The external probes must be connected to the WiFi data logger before
logging into the Cloud for the first time. If an additional probe is to be
connected at a later stage, the WiFi data logger must first be logged
out of the Cloud. The external probe can then be connected and the
WiFi data logger logged in again.
CAUTION
Damage to WiFi data loggers!
- Do not place near any solvents.
- Do not clean using solvents.
CAUTION
Potential damage to the optical surfaces (THL, S-Lux and S-LuxUV)
- Do not use sharp objects.
- Only use soft cleaning cloths.
- Do not use aggressive cleaning agents.
CAUTION
Potential damage to the optical components (IAQ)
- Avoid any vibrations, the factory calibration may be altered. Check the
readings in fresh air 350 to 450 ppm CO
2
(urban air up to 700 ppm CO
2
).
- Prevent condensation. This can result in elevated CO
2
readings.
- Do not use aggressive cleaning agents.
The data loggers must only be mounted vertically. Here, the
connections must point downwards. In the case of data loggers with a
display, you need to pay attention to the reading direction. Otherwise,
the measuring accuracy might be diminished.