7-5.
Setting the Buzzer Type(F4)
This function allows setting the type of the buzzer which sounds at the end of a run.
※
Factory default value is "0".
Set value Buzzer type
0 “Beep” three times
1 “Pip” four times
2 Short “beep” three times
3 Long “beep” three times
4 “Pip-pip” three times
5 “Pip-pip-pip” two times
6 “Pip-pip-pip-pip” three times
7 Two times “pip-pip-pip” and “pip” seven times
7-6.
Setting the Buzzer Time(F5)
This function allows setting the time (the number of seconds) for which the buzzer continues to
sound at the end of a run.
※
Factory default value is "4".
Set value [sec] Description
0 No buzzer sounds.
1
〜
99 Buzzer sounds for 1 to 99 seconds.
7-7.
Setting the Lid-up Monitoring Time(F6)
This function allows checking the open/closed status of the lid and setting the lid-up
monitoring time. If the chamber lid is left open for a long time with the instrument turned on,
the chamber may become frosted inside and cannot be sufciently cooled down. To prevent
frost from forming, the alarm buzzer will sound if the lid remains opened for a time exceeding
the set monitoring time.
※
Factory default value is "18".
(Lid-up monitoring time is 3 minutes.)
Set value [×10sec] Description
0 No lid-up check
1
〜
99 Lid-up monitoring time can be set from 10 seconds to 990 seconds