— 45 —
7-5.
Setting the Buzzer Type(F4)
This function allows setting the type of the buzzer which sounds at the end of a run.
Factory default value is "0".
Set value Buzzer type
0 “Beep” three times
1 “Pip” four times
2 Short “beep” three times
3 Long “beep” three times
4 “Pip-pip” three times
5 “Pip-pip-pip” two times
6 “Pip-pip-pip-pip” three times
7 Two times “pip-pip-pip” and “pip” seven times
7-6.
Setting the Buzzer Time(F5)
This function allows setting the time (the number of seconds) for which the buzzer continues to
sound at the end of a run.
Factory default value is "4".
Set value [sec] Description
0 No buzzer sounds.
1
99 Buzzer sounds for 1 to 99 seconds.
7-7.
Setting the Lid-up Monitoring Time(F6)
This function allows checking the open/closed status of the lid and setting the lid-up
monitoring time. If the chamber lid is left open for a long time with the instrument turned on,
the chamber may become frosted inside and cannot be sufciently cooled down. To prevent
frost from forming, the alarm buzzer will sound if the lid remains opened for a time exceeding
the set monitoring time.
Factory default value is "18".
(Lid-up monitoring time is 3 minutes.)
Set value [×10sec] Description
0 No lid-up check
1
99 Lid-up monitoring time can be set from 10 seconds to 990 seconds