5-3. Mounting/Dismounting the Rotor
CAUTION
Ensure that there is no damage or foreign material on the motor shaft.
Otherwise, the rotor cannot be mounted properly on the motor shaft, thereby
causing the rotor to fall off during spinning.
Ensure that there is no damage, deformation
or corrosion in rotors.
Otherwise, the rotor may break during spinning,
thereby causing damage to the instrument.
Ensure that there is no foreign material in the
rotor or in the chamber before operating the
instrument.
If any, it may break a sample during spinning and
its broken pieces may scatter, resulting in a
serious accident. Or an unbalanced load may
occur, thereby causing the rotor to fall off.
Ensure that the rotor is securely installed.
Otherwise, it may cause the rotor to fall off during spinning.