5-5. Loading Samples
CAUTION
Load the rotor symmetrically around the
motor shaft. opposing adapters, tubes and
samples should be of the same type and
tubes should be lled equally.
Otherwise, it may cause an unbalanced load and
apply excessive force to the rotor, thereby causing
the rotor to fall off.
Ensure that no water drop or foreign material is inside the tube hole.
Otherwise, unbalanced load may occur, applying excessive force to the motor
shaft, thereby causing the rotor to fall off.
Ensure that the load of the tube or sample is appropriately balanced.
Otherwise, it may cause an unbalanced load and apply excessive force to the
motor shaft, thereby causing the rotor to fall off.
Be sure to attach the rotor lid to the rotor if it is provided and make
sure it is surely secured and closed.
Otherwise, it may cause the rotor to fall off during spinning.