— 40 —
The speed can be set by inputting the RCF value.
<
To set speed according to RCF value
>
Please refer to "5. OPERATION" for setting of any parameters other than speed.
CAUTION
Do not exceed the maximum allowable RCF of the tube.
Failure to do so may break the tube and cause the rotor to fall off during
spinning.
When the mass density of sample exceeds 1.2 g/cm
3
, do not perform
operation over calculated allowable speed of the rotor.
An excessive speed may damage the rotor, bucket or rack, thereby causing
the rotor to fall off.
In speed setting mode, press the key to switch the display from speed to
RCF value.
Press the key again. A single "pip" sounds. The Speed/RCF display
begins to ash.
After checking that the “G” unit lamp lights up, turn the Jog Dial to set a desired RCF.
The rpm/G unit lamps light up alternately by pressing the key.
After setting the RCF, leave it for approximately two seconds to nalize the setting.
SPEED/RCF
SPEED/RCF
SPEED/RCF