TH2883 Series Operation Manual Chapter 8 SCPI Command reference
8-13
The :DIFFzone:RANGe command sets the comparison range of the Diffzone comparator.
The :DIFFzone:RANGe? query returns the current comparison range of the Diffzone comparator.
Command syntax: COMParator:DIFFzone:RANGe <start pot, end pot>
Where,
<start pot> start point of comparison range. NR1 format, range from 0 to 6000, without unit.
<end pot> end point of comparison range. NR1 format, range from 0 to 6000, without unit.
For example: WrtCmd(“COMP:DIFF:RANG 0,960 ); set the comparison range from 0 to 960.
NOTE: The end point value should be larger than that of the start point; otherwise an
error message will be displayed on the system message line.
Query syntax: COMP:DIFF:RANG?
Return format: <start pot>, <end pot><NL^END>
start pot, end pot are NR1 format.
The :DIFFzone:DIFFerence command sets the difference limit value of Diffzone comparator.
The :DIFFzone:DIFFerence? query returns the current difference limit value of the Diffzone
comparator.
Command syntax: COMParator:DIFFzone:DIFFerence <value>
Where,
<value> can be NR1, NR2 or NR3 format without unit.
For example: WrtCmd( “COMP:AREA:DIFF 2.5” ); set the difference limit value to 2.5%.
NOTE: Here <value> is a percent value, for example, input 2.5 for 2.5%.
Query syntax: COMParator:DIFFzone:DIFFerence?
Return format: <NR3><NL^END>
The :COROna[:STATe] command sets the Corona comparator to ON or OFF.
The :COROna[:STATe]? query returns the current Corona comparator state.
Command syntax:
ON
COMParator:COROna[:STATe] OFF
1
0
For example: WrtCmd( “COMP:CORONA ON” ); turn on the Corona comparator.
Query syntax: COMParator:COROna[:STATe]?
Return format: <CORONA STAT><NL^END>
Where, < CORONA STAT >is as follows:
ON the Corona comparator function is ON
OFF the Corona omparator function is OFF