6. Basic Operation
28
When the instrument system detects a shock, this
function informs the operator of it.
When the instrument’s installation status (height)
is sharply changed by the contact of the operator
or the like, this function stops auto leveling to
keep the operation accuracy and informs the
operator of the situation. The three lamps blink at
the same time as shown at the right.
After 1 minute has passed since the auto leveling
function was activated and the laser beam was emitted, this function works.
The height alert function does not work in the “Manual” mode.
How to reset
1
Turn off the power switch.
2 Check whether the instrument is installed correctly.
3 Turn on the power switch. Auto leveling starts again. After auto leveling is finished, the
laser beam is emitted.
4 Make sure that the laser beam is set at the correct height. Then, restart the operation.
CHeight Alert ON/OFF(Refer to “Height Alert ON/OFF” (p. 46)
6.2 Height Alert Function
Shock is given to the instrument.
Height Alert Display
(Flashing)