EasyManua.ls Logo

Topcon RL-SV2S - Horizontal Rotation Cone Error Check

Topcon RL-SV2S
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8. Check and Adjusting
51
Horizontal Rotation Cone Error
Perform the following check after completing “Horizontal Calibration” on the previous page.
1 Set up the laser centered between two walls approximately 50 m (164 ft) apart. Orient the
instrument so one axis, either X or Y, is facing the walls. Grade should be set to 0.00% in
both axes.
2 Locate and mark the position of the RL-SV2S beam on both walls using the LS-80L.
3 Turn off the RL-SV2S and move the RL-SV2S closer to wall A (1 m to 2 m /3 ft to 6 ft).
Do not change the axis orientation of the RL-SV2S. Turn the RL-SV2S on.
4 Again locate and mark the position of the RL-SV2S beam on both walls using the LS-80L.
5 Measure the distance between the first and second marks on each wall.
Standard
location
Wall B
Wall A
Approx.µ°í
Wall B
Wall A
Cone Error

Related product manuals