EasyManua.ls Logo

Topcon RT-5SW - Grade Setting Error Check; Checking Grade Setting Accuracy

Topcon RT-5SW
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
Grade Setting Error
Perform the following check only after completing “Horizontal Calibration” and
“Horizontal Rotation Cone Error”.
(1) Checking
1 Securely position two nails or stakes exactly 30m (93 ft) apart. These positions will be called
Nail 1 and Nail 2.
2 Set up the instrument on a tripod approximately 1m to 2m (3 ft to 6 ft) in front of Nail 1 on the
line created by Nail 1 and Nail 2. Position the instrument so the “X-” axis is directly on the line
created by Nail 1 and Nail 2.
3 Turn on the instrument and verify the grade setting is 0.00% in both axes. Turn on the LS-70A
laser sensor and set it to high precision detection. Using the staff, read the elevation of the
laser beam in millimeters at Nail 1 and Nail 2. Designate the elevation at Nail 1 as “h1”, at Nail
2 as “h2”.
Due to the accuracy required to perform this check, the staff used in this procedure must
be graduated in millimeters.
Level
sensor
Staff
Nail 1 Nail 2

Related product manuals