7–
ROD LENGTH
If the gauge depth seems incorrect after calibrating the depth, the Rod
Length function allows the operator to set the rod length manually. The
operator should not need to set the rod length unless a new front panel
assembly is installed or the RAM is corrupt. To set the rod length
manually, press
SHIFT and SPECIAL for:
Press 19 to display:
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press START/ENTER. The display will be:
Press YES for:
To access the Rod Length function, press 6. The gauge provides the
operator with two rod-length options: 8 or 12 inches. The display is
Press 1 or 2 to select the correct rod length. The gauge returns to the
Extended Functions menu. To return to the Ready mode, press NO/CE.
If taking measurements in the automatic indicator mode, the operator
should calibrate the depth strip after manually entering the rod length.
YES- Next menu
4-
C
al
i
b.
Da
te
5-
Ca
lib
.
C
on
st
.
6-
Rod Length
EXTENDED
FUNCTIONS
Enter Code
----
and Press ENTER
SPECIAL FUNCTION
YES- Next menu
1-
STAT TEST
2-
DRIFT TEST
YES- Next menu
1- Time/Date
2-
Customer Name
3-
Se
ria
l
N
um
be
r
Set Rod Length