DRIFT TEST
********************************
Gauge Serial #: 123456
--------------------------------
Average:
3751
0.3 %
Place the gauge on the reference standard block. Make sure the source rod
is in the SAFE position and press
〈START/ENTER〉. The display will be:
After the gauge completes five 4-minute counts, the display will be:
To view the moisture data from the last drift test, press 〈START/ENTER〉.
The display is:
VIEWING DRIFT TEST DATA
The data for each of the readings may be viewed on the screen or may be
printed. This function is the same as viewing the stat test data.
Refer to page 6–5 for instructions on viewing or printing drift test data.
Figure 6–2 shows a sample drift test printout.
Figure 6–2. Drift Test Printout Sample
Avg
cnts: xx
D
rift:
x
.x
%
P
View Drift data?
Avg cnts: xx
D
rift:
x
.x
%
P
-
EN
TE
R
f
or
Mo
is
t
Read #x Time:xxx