6–8
********************************
DRIFT TEST
********************************
Gauge Serial #: 123456
O5/09/2000 3:01 PM
READING |
COUNTS
# | DENSITY MOISTURE
--------------------------------
Average:
3751
Drift:
0.3 %
(PASS)
671
0.2 %
(PASS)
Place the gauge on the reference standard block. Make sure the source rod
is in the SAFE position and press
START/ENTER. The display will be:
After the gauge completes five 4-minute counts, the display will be:
To view the moisture data from the last drift test, press START/ENTER.
The display is:
VIEWING DRIFT TEST DATA
The data for each of the readings may be viewed on the screen or may be
printed. This function is the same as viewing the stat test data.
Refer to page 6–5 for instructions on viewing or printing drift test data.
Figure 62 shows a sample drift test printout.
1
3769
672
2
3801
661
3
3777
670
4
3740
694
5
3669
660
Figure 62. Drift Test Printout Sample
MOIST DRIFT TEST
Avg
cnts: xx
D
rift:
x
.x
%
P
View Drift data?
DENS DRIFT TEST
Avg cnts: xx
D
rift:
x
.x
%
P
-
EN
TE
R
f
or
Mo
is
t
-
DRIFT TEST
-
Read #x Time:xxx