Model 3440-L 6–13
6. SPECIAL FUNCTIONS
DISABLE A SPECIAL CALIBRATION
To disable a previously enabled special calibration, press 2 from the
SPEC CALIB. menu. The display will be:
NORMAL
CALIBRATION
ENABLED!
The gauge will return to the READY display.
PERFORM A SPECIAL CALIBRATION (RECALIBRATION)
Before performing a special calibration, the following steps must be taken:
1. The sample area must be prepared for testing.
2. Gauge readings must be taken on each area.
3. A wet density measurement using an alternative method must be
performed.
To perform a new special calibration, prepare a test site and position the
gauge. Position the rod at the correct depth and press 3 from the SPEC
CALIB. menu. The display will be:
Input rod depth:
(Enter 0 for BS)
xxinches
(Press ENTER)
Press START/ENTER after input of the rod depth. The display is:
SPECIAL CALIB.
Depth: x inches
Time: x min.
Press START

Table of Contents

Other manuals for Troxler 3440

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3440 and is the answer not in the manual?

Troxler 3440 Specifications

General IconGeneral
Model3440
CategoryMeasuring Instruments
DisplayLCD
Power SupplyRechargeable Battery
Measurement TypeDensity and Moisture
Radiation SourceCesium-137 and Americium-241
ApplicationSoil
Depth of MeasurementUp to 12 inches (300 mm)
Measurement Time1 minute (standard)
Battery LifeApproximately 8 hours
Operating Temperature14°F to 122°F (-10°C to 50°C)

Related product manuals