Model 3440
6–7
DRIFT TEST
If the stat test has already been performed (and has passed), but gauge
readings seem to be fluctuating from test to test, the Drift Test can be
performed to check the long-term drift of the 3440.
Pass and Fail limits are averages from the stat test and the difference in the
drift test average. If the average differences exceed 0.5% for density or 1%
for moisture, contact your Troxler service department.
Without turning off or moving the gauge from the time the stat test was
taken, wait 3 to 8 hours and implement the drift test.
NOTE
The gauge should not be turned off between the stat and drift
tests. The stat test must be current. The gauge must not be
moved between the stat and drift tests to eliminate possible
failure due to positioning changes.
To execute the drift test, press SHIFT and SPECIAL for the display:
Press 2 for the display:
EXECUTE DRIFT TEST
To execute the drift test, press NO/CE at the above display:
SPECIAL FUNCTION
YES- Next menu
1-
STAT TEST
2-
DRIFT TEST
-
DRIFT TEST
-
Want to view
l
as
t
Dri
ft
da
ta
?
Put gauge on Std
B
lo
ck
,
r
od
in
SAFE
pos
iti
on
&
press
START
6. SPECIAL FUNCTIONS