CALIBRATION STANDARD
The Calibration Standard function allows the operator or service
technician to input the factory calibration standard count.
To execute Calibration Standard function, press 〈SHIFT〉 and 〈SPECIAL〉
for:
Press 〈19〉 to display:
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press 〈START/ENTER〉. The display will be:
Press 〈YES〉 three times for:
Press 〈1〉 〈0〉 to display:
Input the calibration standard count and press 〈START/ENTER〉.
The display will return to the Extended Functions menu.