Model 3440
7–11
CALIBRATION STANDARD
The Calibration Standard function allows the operator or service
technician to input the factory calibration standard count.
To execute Calibration Standard function, press SHIFT and SPECIAL
for:
Press 19 to display:
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press START/ENTER. The display will be:
Press YES three times for:
Press 1 0 to display:
Input the calibration standard count and press START/ENTER.
The display will return to the Extended Functions menu.
YES- Next menu
10
-
Ca
l
ib
.
S
td
1
1-
R
emo
te
Cn
tr
l
EXTENDED
FUNCTIONS
Enter Code
----
and Press ENTER
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
YES- Next menu
1- Time/Date
2-
Customer Name
3-
Se
ria
l
N
um
be
r
C
al
ib
.
S
tan
da
rd
Co
un
t:
x
I
npu
t
a
nd
Press ENTER
7. EXTENDED FUNCTIONS