Model 3440
2–7
ASPHALT MODE, % MARSHALL. If % Marshall (or % compaction
compared to a Marshall target) is required, press
1 from the last display
on page
2–6.
% Marshall = (WD/Marshall)
×
100
To enable % Voids, press YES.
% VOIDS = 100
×
[1 (WD/Voidless)]
NOTE
For the gauge to calculate % Voids, a voidless density must be
entered into the gauge as described on page 5–15. The voidless
density is the maximum theoretical density, or Rice test result.
After a brief delay, the screen display returns to the READY display.
ASPHALT MODE, 100 %MA. If 100 %MA is required, press 2
from the last display on page 2–6.
NOTE
The term % Voids is defined as 100 %Voidless.
100 %MA = 100
×
[1 (WD/Marshall)]
To enable % Voids, press YES.
%VOIDS = 100
×
[1 (WD/Voidless)]
After a short delay, the display returns to READY. Chapter 4 provides
instructions for taking measurements in Asphalt mode.
ASPH
AL
T
:
%
MA
Do you want
to
ENABLE %VOIDS
also?
ASPH
AL
T
:
%
MA
%
VO
I
DS
ASPHALT: 100-%MA
Do you want
to
ENABLE %VOIDS
also?