Press ENTER. The compactor stores the new calibration
constants based on the Measured and Standard values. It then
displays the message Calibration Finished and lowers the
ram to its home position. When the ram reaches the home
position, the compactor returns to the Calibration menu shown
on page 4–5. Open the chamber door and remove the height
standard assembly, pucks, and specimen papers.
ANGLE OFFSET
The Model 5850v2 can be set to a gyration angle of 0.00 to 1.50
degrees. As noted earlier, Troxler recommends verifying the angle
after every 800 samples or once per year using an internal angle
device, such as the Dynamic Angle Verification (DAV) device
available from Troxler. Refer to page 4–3 for instructions on
verifying the angle.
Use the internal angle device during verification to take two bottom
angle measurements and two top angle measurements. Determine
the average of these four measurements. If the average angle is more
than ±0.03° from the target, the Angle Offset function can be used to
compensate for the difference by comparing control system
feedback parameters to the average measured internal angle.
To use the Angle Offset function, press 3 at the Calibration menu
shown on page 4–5. The compactor displays:
-Angle Offset-
0.00 Degrees
1- Change
2- Enable