B. Bline Mode
Press the and buttons to scroll the cursor to the sensor type. Press the
button to enter the black-mark sensor calibration mode.
B-1 Automatic
When enter the [Automatic] option, you will see following message and printer will
feed the black mark label to calibrate the sensor sensitivity automatically. When
calibration process is completed, the LCD screen will return to the previous menu.
B-2 Manual
In case “Automatic” sensor calibration cannot apply to the media, please use
“Manual” function to calibrate the bline sensor manually.
When enter [Manual] option, you will see following message. Please complete there
steps:
1. Press the button to move the
cursor from left digit to right digit, and press
the button to set the value from “0” to
“9” and the “dot/ mm/ inch”. Press the
button to set the paper length into
the printer.
2. Press the button to move the
cursor from left digit to right digit, and press
the button to set the value from “0” to
“9” and the “dot/ mm/ inch”. Press the
button to set the bline size into the
printer.