Page 22 of 63 Pages
12. The POWER light is lit during the entire sterilization period, indicating
connection to the power. The HEAT indicator signifies the operation of
the pressure switch (B10), which keeps the temperature stable by
connecting and disconnecting the heating elements. When the timer
reaches 0, the sterilization process is complete and a buzzer will indicate
that the cycle is complete.
13. If a cycle with drying is required proceed according to para. 14.
If a cycle without drying is required proceed according to para. 15
14. Cycle with drying
14.1 When the pressure gauge reaches 0,
release the locking device and open
the door slightly, in order to permit the
steam to evacuate
14.2 Set the timer on 20-30minutes. Drying
indicator lamp will light.
14.3 When the time is over, turn the
multi-purpose valve knob to "0"
position.
14.4 Open the door and Remove the load.
Warning
Multi-purpose valve knob should be
turned in a clockwise direction only.
15. Cycle without drying
15.1 Turn immediately, the multi-purpose valve knob to EXH.+DRY
(Exhaust and Dry) position.
15.2 When the pressure gauge reaches 0, turn the multi-purpose valve
knob to "0"
15.3 Open the door to remove the load.
Warning
Multi-purpose valve knob should be turned in a clockwise
direction only.
Note:
The sterility of instruments processed in unwrapped cycles
cannot be maintained if exposed to a non-sterile environment.